Abstract

A single-channel phoswich well detector has been assessed and analysed in order to improve beta–gamma coincidence measurement sensitivity of 131mXe and 133mXe. This newly designed phoswich well detector consists of a plastic cell (BC-404) embedded in a CsI(Tl) crystal coupled to a photomultiplier tube (PMT). It can be used to distinguish 30.0-keV X-ray signals of 131mXe and 133mXe using their unique coincidence signatures between the conversion electrons (CEs) and the 30.0-keV X-rays. The optimum coincidence efficiency signal depends on the energy resolutions of the two CE peaks, which could be affected by relative positions of the plastic cell to the CsI(Tl) because the embedded plastic cell would interrupt scintillation light path from the CsI(Tl) crystal to the PMT. In this study, several relative positions between the embedded plastic cell and the CsI(Tl) crystal have been evaluated using Monte Carlo modeling for its effects on coincidence detection efficiency and X-ray and CE energy resolutions. The results indicate that the energy resolution and beta–gamma coincidence counting efficiency of X-ray and CE depend significantly on the plastic cell locations inside the CsI(Tl). The degraded X-ray and CE peak energy resolutions due to light collection efficiency deterioration by the embedded cell can be minimised. The optimum of CE and X-ray energy resolution, beta–gamma coincidence efficiency as well as the ease of manufacturing could be achieved by varying the embedded plastic cell positions inside the CsI(Tl) and consequently setting the most efficient geometry.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , ,