Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1877980 | Applied Radiation and Isotopes | 2010 | 5 Pages |
Abstract
Both 1 and 6Â MBq ampoules were assayed by gamma-spectrometry using HPGe and NaI(Tl) detectors. This method resulted in an activity value with a smaller uncertainty than the primary method. Activity values for the three methods employed were consistent within the uncertainty of measurement.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
H.C. Phillips, L.C. Johansson, J.P. Sephton,