Article ID Journal Published Year Pages File Type
1878003 Applied Radiation and Isotopes 2010 4 Pages PDF
Abstract

A MCNP model was developed for the efficiency calibration of an in situ gamma ray spectrometry system based on a high purity germanium (HPGe) detector. The detector active crystal volume was adjusted semi-empirically against experimental measurements. Calculated full energy peak efficiency curves, over the photon energy range between 50 keV and 5 MeV, are presented for surface and slab source configurations. The effect of different collimator apertures and the contribution of different HPGe crystal regions in the detector response are also shown.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation