Abstract

•Large (9 L) solid Ge-68 epoxy phantom was designed and calibrated.•Activity concentration was calibrated with standard uncertainty of about 1.1%.•Inhomogeneity of Ge-68 activity distribution was estimated at about 0.3%.•Data from PET–CT scans of phantoms were consistent with data from calibration.

We have developed a methodology to calibrate the 68Ge activity concentration in large (9 L) cylindrical epoxy phantoms in a way that is traceable to national standards. The method was tested on two prototype cylindrical phantoms that are being used in a clinical trial and gave 68Ge activity concentration values with combined standard uncertainties of about 1.1%. Imaging data from the phantoms using a calibrated PET–CT scanner gave values consistent with the calibrated activity concentrations within experimental uncertainties.