The work accomplished within the participation at the 2012 key comparison of Tc-99 is presented. The solution was standardized for the first time in IFIN-HH by two methods: LSC-TDCR and 4Ï(PC)β-γ efficiency tracer. The methods are described and the results are compared. For the LSC-TDCR method, the program TDCR07c, written and provided by P. Cassette, was used for processing the measurement data. The results are 2.1% higher than when applying the TDCR06b program; the higher value, calculated with the software TDCR07c, was used for reporting the final result in the comparison. The tracer used for the 4Ï(PC)β-γ efficiency tracer method was a standard 60Co solution. The sources were prepared from the mixture 60Co+99Tc solution and a general extrapolation curve, type: NβTcâ99/(M)Tcâ99=f [1âεCoâ60], was drawn. This value was not used for the final result of the comparison. The difference between the values of activity concentration obtained by the two methods was within the limit of the combined standard uncertainty of the difference of these two results.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
M. Sahagia, A. Antohe, R. Ioan, A. Luca, C. Ivan,