Abstract
In this work we compare the kVp estimate between CDTN multipurpose instrument, UnforsXI and Radcal 4075 meters under different combinations of voltage and filtration. The non-invasively measurements made using x-ray diagnostic and interventional radiology devices show similar tendencies to increase the kVp estimate when aluminum filters are placed in the path of the x-ray beam. The results reveal that the kVp estimate made by the CDTN multipurpose instrument is always satisfactory for highly filtered beam intensities.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
A.T. Baptista Neto, B.B. Oliveira, L.O. Faria,