Article ID Journal Published Year Pages File Type
1878987 Applied Radiation and Isotopes 2015 5 Pages PDF
Abstract
In this work we compare the kVp estimate between CDTN multipurpose instrument, UnforsXI and Radcal 4075 meters under different combinations of voltage and filtration. The non-invasively measurements made using x-ray diagnostic and interventional radiology devices show similar tendencies to increase the kVp estimate when aluminum filters are placed in the path of the x-ray beam. The results reveal that the kVp estimate made by the CDTN multipurpose instrument is always satisfactory for highly filtered beam intensities.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , ,