Abstract

Pulses from detectors used for radioactivity measurement can vary in size by several orders of magnitude. Large pulses will lead to saturation at the preamplifier output and extension of the pulse length. As a consequence, the dead time of the system increases and pulses may be lost. Electronic design techniques employed to protect against overloading tend to increase the amplifier noise level. However, an optoelectronic method of overload protection has been devised which has only a negligible effect on noise. An infrared light emitting diode interfaced to the output of the preamplifier is linked by fibre optic cable to an ultra-low leakage photodiode at the input. The conduction of the photodiode increases with the amplitude of the preamplifier output signal. Excess current is thereby prevented from entering the preamplifier and causing saturation. The preamplifier has been tested on 4π beta–gamma and gas counting systems and found to give good protection against overloading.

► A preamplifier for radioactivity measurements has been developed. ► Low noise. ► Current sensitive. ► Optoelectronic overload protection.