Abstract

A new triple-to-double coincidence ratio (TDCR) system has been established at CMI as an additional technique for primary activity standardisation. Details of the detectors and the electronics are given. Construction of the optical chamber was validated using measurement of a 3H standard—the efficiency achieved with Ultima Gold was 58%. Several methods of efficiency variation (coloured bands painted on vial, positioning and photomultipliers defocussing) are compared. Activity concentrations of 204Tl and 45Ca were determined with the TDCR method and with the efficiency tracer technique. Advantages of the TDCR method compared to the efficiency tracing method and agreement between results are shown.

► The new TDCR system was constructed and verified at CMI. ► Maximum detection efficiency was 58% for H-3. ► Consistent results for Tl-204, Ca-45 with ∼50% lower uncertainty than the tracer method.