Abstract

A simple modification of the GE Tracerlab FX C Pro system which enabled performance of both solvent capture (loop method) and conventional solution phase [11C]methylation in the same module is described. By the quick setup and automated method, [11C]carfentanil and [11C]raclopride could be prepared in rapid succession without opening the hot cell. The radiochemical yields were over 40% and 30% (decay-corrected and based on [11C]methyl iodide) for [11C]carfentanil and [11C]raclopride, respectively. The radiochemical purities were greater than 95% and specific activities over 5 Ci/mmol for both tracers. The modification is extremely easy and can be utilized for multiple syntheses of other 11C-labeled radiopharmaceuticals in a fast and reliable manner.