Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1879590 | Applied Radiation and Isotopes | 2008 | 7 Pages |
Abstract
Measurements of 241Am L X-ray emission probabilities were conducted using both HPGe and Si(Li) detectors. The efficiency calibrations of these detectors were performed by means of a tunable, monochromatic photon beam and the determination of the thickness of absorbing materials inside the detector. These efficiency calibrations were obtained without any reference to radionuclide decay data, and with 0.8% relative standard uncertainty. The complex L X-ray region was processed using Voigt functions and by taking account of the detector response function established with the monochromatic beam. Twenty-two components of the L X-ray group were identified and quantified. The present results are compared with previously published data.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
M.C. Lépy, J. Plagnard, L. Ferreux,