Article ID Journal Published Year Pages File Type
1879641 Applied Radiation and Isotopes 2008 5 Pages PDF
Abstract
The two-dimensional extrapolation technique and dead time extrapolation technique of 4πβ-γ-coincidence counting were used to standardize the activity of 67Ga. The counting results of the two absolute counting techniques showed good agreement while the two-dimensional extrapolation technique could save about 50% of the total counting time in this study. The accuracy of the Capintec CRC-15R radionuclide calibrator used by the radiopharmacy of INER was studied in this research. A new calibration setting number, 99, was recommended to the radiopharmacy of INER in the Bayer 10 mL sterile empty vial geometry.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , ,