Abstract
Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Menglong M.D., Liangan B.S., Shuyu B.S., Guangfu B.S.,