1880507 Radiation Measurements 2015 6 Pages PDF
Abstract

•The compact radiation sensor consists of a CsI(Tl) crystal and a typical SiPM.•The optimal parameters of the crystal are a tapered length and a total length.•The parameters are determined by comparing the angular response and the FOM.

In this study, a compact radiation sensor for integration into mobile applications is proposed consisting of a CsI(Tl) crystal and a silicon photomultiplier with a 3 × 3 mm2 active area. Considering the size limit of mobile devices, the optimal size of the CsI(Tl) crystal was determined so that it has the best figure of merit (the absolute detection efficiency over the relative energy resolution) in the gamma energy spectrum while satisfying the criteria for the uniformity of the angular response. The final suggested dimensions consist of a cylindrically shaped crystal with a diameter of 5 mm, a total length of 5 mm and a tapered head length of 1 mm.