Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1880523 | Physica Medica | 2013 | 5 Pages |
Abstract
Radiation leakage of up to 170% of the prescribed dose has been found at the back surface of these applicators. Although this side is not usually directed to the patient, in some applications such as the treatment of a lesion on the nose, special care must be taken to avoid unexpected and unnecessary irradiation of the eyes. A possible solution could be to add additional shielding to the applicator in order to reduce this leakage or to put some shielding to protect the eyes. Additionally, a new concept design of the Valencia applicators using more shielding material in the applicator backside is proposed.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
D. Granero, J. Perez-Calatayud, F. Ballester, Z. Ouhib,