Article ID Journal Published Year Pages File Type
1880601 Radiation Measurements 2014 4 Pages PDF
Abstract

•A simple brachytherapy QA phantom has been instrumented.•Dwell positions have been estimated according to measured dose rate by GaN probes.•Measured dose rate shows 1% average deviation from calculated results.•Determined dwell position precision and measureable range was studied.

For High Dose Rate (HDR) brachytherapy Quality Assurance (QA), we propose a seed position verification method by using two GaN dosimetric probes integrated in an instrumented phantom. As dose rate measured from one probe is related to the seed distance, seed position inside catheter can be determined. The proposed two-probe method extends measureable range and improves accuracy in comparison with the use of one single probe. One instrumented phantom implementing the method was designed, fabricated and tested in clinical conditions. The measureable range of the two-probe method is 105 mm compared to 36 mm using single probe. The validity of the method was experimentally verified.

Keywords
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , ,