Article ID Journal Published Year Pages File Type
1880630 Radiation Measurements 2014 4 Pages PDF
Abstract
To develop an improved dosemeter to assess photon and beta exposures of the eye lens, and in response to issues surrounding the preferred values of Hlens to be used for guiding operational radiation protection, a programme of re-optimization of the current PHE thermoluminescence dosemeter has been performed. In particular, refinements of the filter located in front of the sensitive 7LiF:Mg,Cu,P element have been considered, so that the dose response characteristics of the device provide a better and more conservative estimate of risk. The investigation was performed using the Monte Carlo modelling software MCNP5, to produce a final design that featured a filter containing a 9.5 mm diameter polypropylene hemisphere truncated to a maximum thickness of 3.0 mm. The responses of this design in photon and electron fields are presented here, contrasted against those of the existing PHE eye dosemeter, with respect to the operational quantity Hp(3,E,θ) and both current and suggested values for the absorbed dose per fluence risk profile for the lens of the eye.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation