Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1880632 | Radiation Measurements | 2014 | 4 Pages |
•Dosimetry using OSL of BeO.•Environmental dosimeter with modified personal dosimeter and linear algorithm.•Use of doped BeO chips for better energy response.
In the course of this work, the possibility of the measurement of ambient dose equivalent H*(10) with the BeOSL dosimetry system was evaluated. Calculations for the energy response of the 2-element BeOSL dosimeter for irradiation with H*(10) were performed. The response doesn't fulfil the requirements of IEC 62387-1. Especially the response for photon energies of 60–100 keV is to low. It is possible to correct this under response using a modified BeOSL 2-element dosimeter and a linear algorithm. So the national requirements for a H*(10) dosimeter in Germany can be fulfilled. An incidence angle independent measurement is not possible because for several angles of incidences (>60°) the filters of the 2-element dosimeter doesn't shield the correct element. Another material which is more suitable for the H*(10) measurement was tested. So the doping of BeO with Lanthanum leads to an enhanced energy response for measurement of H*(10). Furthermore a higher OSL sensitivity was found for this material. Further tests on the influence of Lanthanum concentration on the dosimetric properties are necessary.