Abstract
The results showed the differences in the average TL-efficiency for these two TL-materials. For LiF:Mg,Ti the relative efficiency is within a few percent from unity for any of the analyzed values of shielding, altitude and solar cycle conditions. This means that one can assume cosmic radiation doses measured in Low Earth Orbit (LEO) with LiF:Mg,Ti detectors to be correct within such uncertainty. LiF:Mg,Cu,P underestimates the cosmic radiation doses by more than 15% in all cases. Altitude and solar cycle were found to have a very weak influence on the TL efficiency. In contrast, the influence of shielding thickness is quite significant. The reason for this is a change of contributions of radiation field components: trapped protons dominate at low shielding (97% of dose at 1 g/cm2), but are negligible above 60 g/cm2, as well as changes within GCR spectrum (increase of dose due to lower LET secondaries for higher shielding). Shielding thickness affects both TLD types in different ways: the efficiency of LiF:Mg,Cu,P increases with increasing shielding thickness, while the efficiency of LiF:Mg,Ti shows some fluctuations, with a weak minimum for 60 g/cm2. The response ratio of these TLDs decreases monotonically with the shielding thickness and could be used as an indicator for the average shielding conditions in which the TLDs were exposed.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation