Abstract
The maximum energy of the 241Am-Be neutron source was estimated as 10.9 ± 3.0 MeV. The consistency between the determined energy and the other reported values confirms the validity of using the average stopping power in the superheated drop detectors. The average stopping power was also used to determine the threshold neutron energy as a function of external applied pressure at different temperatures. Knowing the threshold neutron energy as function of applied pressure, can be used in pressure scanning method for neutron spectrometry by superheated drop detectors.