Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1882167 | Medical Dosimetry | 2007 | 8 Pages |
Abstract
Stereotactic radiosurgery and image-guided radiotherapy (IGRT) place enhanced demands on treatment delivery machines. In this study, we describe a high-dose-rate output accelerator as a part of our stereotactic IGRT delivery system. The linac is a Siemens Oncor without a flattening filter, and enables dose rates to reach 1000 monitor units (MUs) per minute. Even at this high-dose-rate, the linac dosimetry system remains robust; constancy, linearity, and beam energy remain within 1% for 3 to 1000 MU. Dose profiles for larger field sizes are not flat, but they are radially symmetric and, as such, able to be modeled by a treatment planning system. Target localization is performed via optical guidance utilizing a 3-dimensional (3D) ultrasound probe coupled to an array of 4 infrared light-emitting diodes. These diodes are identified by a fixed infrared camera system that determines diode position and, by extension, all objects imaged in the room coordinate system. This system provides sub-millimeter localization accuracy for cranial applications and better than 1.5 mm for extracranial applications. Because stereotactic IGRT can require significantly longer times for treatment delivery, the advantages of the high-dose-rate design and its direct impact on IGRT are discussed.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
John E. Ph.D., Heather S. M.D., Mark C. M.D., Edward C. M.S., Kathleen M. C.M.D., Timothy C. M.D., John M. M.D.,