Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1883685 | Physica Medica | 2012 | 10 Pages |
A method to carry out stoichiometric calibrations of CT scanners employed in radiotherapy treatment planning is proposed. The method is based on a simple parametrization of the CT number of a substance, which involves only two variables to describe the substance (electron density and one effective atomic number) and one parameter to describe the beam. The method was tested experimentally on a group of beams. A set of no tissue-like substances of known densities and elemental compositions were employed as calibrators. CT number-to-density curves (RED curves) were calculated with the proposed parametrization and compared to those measured with a commercial density phantom. Differences between the electron densities assigned by the calculated RED curves and the measured ones were in the range 0.009–0.019 (RMS). The proposed method may be employed to carry out accurate stoichiometric calibrations by using only one suitable substance as calibrator, not necessarily tissue-like.