Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1884745 | Radiation Measurements | 2006 | 7 Pages |
The most common ionization chamber used in computed tomography dosimetry is the “pencil ionization chamber”. It is a special cylindrical dosimeter developed for attending computed tomography beams particularities. In this study, a Victoreen pencil ionization chamber was submitted to a set of tests for a detailed evaluation of its operational characteristics. Such as many kinds of detectors, especially field instruments, this ionization chamber had originally a preamplifier to keep it electrically more stable. In this study, the performance of the chamber was analyzed with the original preamplifier and after its removal, and the results were compared. The objective of the preamplifier removal was to enable connecting the chamber to other kinds of electrometers available in laboratories. The behavior of the pencil ionization chamber before and after the removal of the preamplifier was very similar, and the results obtained were always within the limits of international recommendations. The results obtained in both situations allow, if necessary, the preamplifier removal of the system without lack of precision in the measurements.