Article ID Journal Published Year Pages File Type
1885716 Radiation Physics and Chemistry 2017 9 Pages PDF
Abstract

•The supralinearity of GFF and TLD-100 was highly dependent on the UHD.•The maximum supralinearity f(D)max of TLD-100, occurs around 10 kGy.•TL kinetic model can be used to explain the glow peak behavior of doses

The thermoluminescence (TL) response of Germanium Flat Fiber (GFF) and TLD-100 irradiated with 2.5 MeV electrons for the doses up to 1 MGy were studied and compared. The aim was to evaluate the TL supralinearity response at an ultra-high dose (UHD) range and to investigate the change in kinetic parameters of the glow peaks, as the doses increases up to 1 MGy. It is found that the critical dose limit (CDL) of GFF is 5 times higher as compared to TLD-100. CDL is determined by the dose at the maximum supralinearity, f(D)max. It is also found that annealing the TLD-100 and GFF with temperature more than 400 °C is required to reset it back to its original condition, following radiation doses up to 1 MGy. It is also noticed the strange behavior of Peak 4 (TLD-100), which tends to be invisible at the lower dose (<10 kGy) and starts to be appeared at the critical dose limit of 10 kGy. This result might be an important clue to understand the behavior of TLD-100 at extremely high dose range. For both samples, it is observed that the TL intensity is not saturated within the UHD range studied.

Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , ,