Article ID Journal Published Year Pages File Type
1887893 Physica Medica 2007 8 Pages PDF
Abstract
In the comparison the lowest correlation with the standard technique film (X-ray 1) was seen in the purely dose-reduced X-ray. In the range of 1.6-3.4 Lp/mm, both SNR curves have an identical course. Despite a 52% dose reduction in the re-exposed image, both densitometry curves of the conventional and re-exposed X-ray show an almost identical distribution of the transmittance levels. In conclusion film sensitisation provides a technically simple and inexpensive procedure, which is easily integrated into previous film development processes and considerably reduces the patient radiation exposure as well as clearly improving the image quality.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Radiation
Authors
, , , , , ,