Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1889756 | Zeitschrift für Medizinische Physik | 2010 | 10 Pages |
Abstract
We have shown that the standard deviations of the measured air kerma rate for the Co-60 source are generally larger than those of the Ir-192 source. The measurements with the well chamber had the lowest deviation from the certificate value. In all protocols and methods the deviations stood for both nuclides by a maximum of about 1.2% for Ir-192 and 2.5% for Co-60-sources respectively.
Related Topics
Physical Sciences and Engineering
Engineering
Biomedical Engineering
Authors
Golam Abu Zakaria, Wilhelm Schütte, Hasin Anupama Azhari,