Article ID Journal Published Year Pages File Type
2034450 Biologicals 2010 8 Pages PDF
Abstract

Endotoxin contamination is a serious threat to the safety of parenteral drugs, and the rabbit pyrogen test has played a crucial role in controlling this contamination. Although the highly sensitive endotoxin test has replaced the pyrogen test for various pharmaceuticals, the pyrogen test is still implemented as the control test for most blood products in Japan. We examined the applicability of the endotoxin test to blood products for reliable detection and quantification of endotoxin. Nineteen types of blood products were tested for interfering factors based on spike/recovery of endotoxin by using 2 types of endotoxin-specific lysate reagents for photometric techniques. Interfering effects on the endotoxin test by the products could be eliminated by diluting from 1/2 to 1/16, with the exception of antithrombin III. However, conventional lysate reagents that also react with non-pyrogenic substances, such as (1–3)-β-d-glucan, produced results that were not relevant to endotoxin content or pyrogenicity. Our results showed that the endotoxin test would be applicable to most blood products if used with appropriate endotoxin-specific lysate reagents.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Biochemistry, Genetics and Molecular Biology (General)
Authors
, , , , , , , ,