Article ID Journal Published Year Pages File Type
215764 The Journal of Chemical Thermodynamics 2013 10 Pages PDF
Abstract

Benchtop measurement systems have emerged as powerful tools in the ongoing quest for thermophysical property data. We demonstrate that these instruments can yield results of high quality if operated in an informed manner. The importance of sample purity, reproducibility over repeatability, expanded calibration and adjustment protocols, and rigorous uncertainty estimates are emphasized. We report measurement results at ambient atmospheric pressure and temperatures from 343 K to 278 K, including expanded uncertainty estimates, for the density and speed of sound of isooctane and for the speed of sound of toluene. These data are useful for validating the performance of such instruments.

► Detail important considerations for reference quality measurements of thermophysical property data with benchtop instruments. ► Density and speed of sound of isooctane and speed of sound of toluene at (278 K to 343 K) and atmospheric pressure. ► Experimental data compared to available literature data and equations of state.

Related Topics
Physical Sciences and Engineering Chemical Engineering Chemical Engineering (General)
Authors
, , , ,