Article ID Journal Published Year Pages File Type
2159833 Radiotherapy and Oncology 2009 7 Pages PDF
Abstract

Background and purposeTo confirm the feasibility of accelerator-based BNCT (AB-BNCT) for treatment of multiple liver tumors and malignant pleural mesothelioma (MPM), we compared dose distribution and irradiation time between AB-BNCT and reactor-based BNCT (RB-BNCT).Material and methodsWe constructed treatment plans for AB-BNCT and RB-BNCT of four multiple liver tumors and six MPM. The neutron beam data on RB-BNCT were those from the research reactor at Kyoto University Research Reactor Institute (KURRI). The irradiation time and dose–volume histogram data were assessed for each BNCT system.ResultsIn BNCT for multiple liver tumors, when the 5 Gy-Eq dose was delivered as the mean dose to the healthy liver tissues, the mean dose delivered to the liver tumors by AB-BNCT and RB-BNCT was 68.1 and 65.1 Gy-Eq, respectively. In BNCT for MPM, when the mean lung dose to the normal ipsilateral lung was 5 Gy-Eq, the mean dose delivered to the MPM tumor by AB-BNCT and RB-BNCT was 20.2 and 19.9 Gy-Eq, respectively. Dose distribution analysis revealed that AB-BNCT is superior to RB-BNCT for treatment of deep-seated tumors.ConclusionsThe feasibility of the AB-BNCT system constructed at our institute was confirmed from a clinical viewpoint in BNCT for multiple liver tumors and MPM.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Cancer Research
Authors
, , , , , , , , , , , , ,