Article ID Journal Published Year Pages File Type
2161064 Radiotherapy and Oncology 2007 8 Pages PDF
Abstract

PurposeTo investigate the set-up errors and deformation associated with daily placement of endorectal balloons in prostate radiotherapy.Materials and methodsEndorectal balloons were placed daily in 20 prostate cancer patients undergoing radiotherapy. Electronic portal images (EPIs) were collected weekly from anterior–posterior (AP) and lateral views. The EPIs were compared with digitally reconstructed radiographs from computed tomography scans obtained during pretreatment period to estimate displacements. The interfraction deformation of balloon was estimated with variations in diameter in three orthogonal directions throughout the treatment course.ResultsA total of 154 EPIs were evaluated. The mean displacements of balloon relative to bony landmark were 1.8 mm in superior–inferior (SI), 1.3 mm in AP, and 0.1 mm in left–right (LR) directions. The systematic errors in SI, AP, and LR directions were 3.3 mm, 4.9 mm, and 4.0 mm, respectively. The random (interfraction) displacements, relative to either bony landmarks or treatment isocenter, were larger in SI direction (4.5 mm and 4.5 mm), than in AP (3.9 mm and 4.4 mm) and LR directions (3.0 mm and 3.0 mm). The random errors of treatment isocenter to bony landmark were 2.3 mm, 3.2 mm, and 2.6 mm in SI, AP, and LR directions, respectively. Over the treatment course, balloon deformations of 2.8 mm, 2.5 mm, and 2.6 mm occurred in SI, AP, and LR directions, respectively. The coefficient of variance of deformation was 7.9%, 4.9%, and 4.9% in these directions.ConclusionsLarger interfractional displacement and the most prominent interfractional deformation of endorectal balloon were both in SI direction.

Related Topics
Life Sciences Biochemistry, Genetics and Molecular Biology Cancer Research
Authors
, , , , , ,