Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
2495601 | Pharmaceutical Methods | 2012 | 5 Pages |
Abstract
Introduction: A simple, precise, and accurate HPLC method for simultaneous estimation of metformin hydrochloride (MET), pioglitazone hydrochloride (PIO), and glimepiride (GLIMP) was developed and validated. Materials and Methods: Chromatographic separation of the drugs was performed by using a Phenomenex-ODS-3 (C-18) column (250 à 4.60 mm, 5 μm) with a mobile phase consisting of methanol:acetonitrile:15 mM potassium dihydrogen phosphate (pH 4) in the proportion of 40:35:25 (v/v) at a flow rate of 1 ml/min. Detection was carried out using a UV-SPD-10AVP detector at 240 nm. Results: The retention time for MET, PIO, and GLIMP were 2.85 ± 0.03 min, 4.52 ± 0.03 min, and 7.08 ± 0.02min, respectively. Parameters such as linearity (0.2-50 μg/ml for MET, 0.2-30 μg/ml for PIO, and GLIMP, respectively), precision (intra-day % RSD was 1.01-3.24 and inter-day % RSD was 1.54-4.09 for MET; intra-day % RSD was 1.03-2.09 and inter-day % RSD was 2.26-3.10 for PIO; and intra-day% RSD was 1.00-3.15 and inter-day % RSD was 1.58-3.07 for GLIMP), accuracy (99.66 ± 0.14 for MET, 98.46 ± 0.40 for PIO, and 98.62 ± 0.39 for GLIMP), specificity and robustness were calculated in accordance with ICH guidelines. Conclusions: The method was proved to be simple, rapid, precise, accurate, and cost effective.
Related Topics
Life Sciences
Biochemistry, Genetics and Molecular Biology
Molecular Medicine
Authors
Vinay Pandit, Roopa S. Pai, Gurinder Singh, Kshama Devi, Satya Narayana, Sarasija Suresh,