Article ID Journal Published Year Pages File Type
2648037 European Journal of Oncology Nursing 2009 5 Pages PDF
Abstract

BackgroundThis study assessed the agreement between infrared tympanic membrane (TM), axillary, corrected axillary (+0.5 °C), oral, and corrected oral (+0.3 °C) to rectal thermometry as reference standard in neutropenic adults. The sensitivity and specificity of the mentioned thermometries in detecting rectal fever (≥38 °C) were also analysed.MethodThis is a comparative diagnostic test study. A total of 400 sets of blinded simultaneous temperature readings were measured from 21 haemato-oncology in-patients with neutropenia following chemotherapy. Three-hundred sets were then randomly sampled. Agreements were analysed using random two-way intraclass correlation (ICC). Sensitivity and specificity were analysed using contingency 2 × 2 table.FindingsBoth right and left TM thermometry have good correlation with rectal thermometry; 0.810 (95% CI, 0.748–0.855) and 0.770 (95% CI, 0.713–0.815) respectively. Axilla thermometry has weak agreement (ICC 0.486 (95% CI, 0.118–0.689)) with rectal thermometry. The sensitivity (sn) and specificity (sp) in detecting rectal fever (≥38 °C) were: right TM (sn) 0.712 (95% CI, 0.586–0.814), (sp) 0.957 (95% CI, 0.920–0.978); oral (sn) 0.561 (95% CI, 0.433–0.681), (sp) 0.983 (95% CI, 0.954–0.995); and axilla (sn) 0.348 (95% CI, 0.238–0.477), (sp) 0.996 (95% CI, 0.973–0.999).InterpretationSingle tympanic membrane thermometry is in good agreement with rectal thermometry. It is more sensitive than oral or axillary thermometry in detecting rectal fever.

Related Topics
Health Sciences Medicine and Dentistry Oncology
Authors
, , ,