Abstract

PurposeTo assess the accuracy of intraocular pressure (IOP) measurements using rebound tonometry over disposable hydrogel (etafilcon A) and silicone hydrogel (senofilcon A) contact lenses (CLs) of different powers.MethodsThe experimental group comprised 36 subjects (19 male, 17 female). IOP measurements were undertaken on the subject's right eyes in random order using a rebound tonometer (ICare). The CLs had powers of +2.00D, −2.00D and −6.00D. Six measurements were taken over each contact lens and also before and after the CLs had been worn.ResultsA good correlation was found between IOP measurements with and without CLs (all r ≥ 0.80; p < 0.05). Bland Altman plots did not show any significant trend in the difference in IOP readings with and without CLs as a function of IOP value. A two-way ANOVA revealed a significant effect of material and power (p < 0.01) but no interaction. All the comparisons between the measurements without CLs and with hydrogel CLs were significant (p < 0.01). The comparisons with silicone hydrogel CLs were not significant.ConclusionsRebound tonometry can be reliably performed over silicone hydrogel CLs. With hydrogel CLs, the measurements were lower than those without CLs. However, despite the fact that these differences were statistically significant, their clinical significance was minimal.