Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
2717308 | The Journal of Foot and Ankle Surgery | 2006 | 5 Pages |
Abstract
The effect of shear stress on a fixated distal syndesmosis of the ankle was evaluated with a servohydraulic materials–testing machine. Eighteen syndesmoses were fixated in a quadricortical technique using 3.5-mm cortical and 4.5-mm cortical stainless steel screws. A shear stress was applied via an axial load in an attempt to simulate weightbearing. The 4.5-mm quadricortical screws produced a higher yield load and peak load (484.3 ± 93.8 N and 597.7 ± 81.4 N) when compared with the 3.5-mm quadricortical syndesmotic screws (412.8 ± 55 N, P = .033 and 511.2 ± 64.4 N). These findings suggest that a larger diameter screw provides greater resistance to an applied shear stress at the distal syndesmosis.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Matthew Hansen, Long Le, Stuart Wertheimer, Eric Meyer, Roger Haut,