Article ID Journal Published Year Pages File Type
2733924 Diagnostic and Interventional Imaging 2012 10 Pages PDF
Abstract

PurposeTo define the optimal acquisition parameters for diffusion tensor imaging of the median nerve at the wrist and to measure the diffusion values.Patients and methodsThe wrists of 14 healthy volunteers were explored using 3T MRI with a dedicated wrist coil. Each examination consisted of six sequences, varying the number of directions (15–20–30) and the b values (500–800–1000–2000 s/mm2). The tractography images were evaluated by two radiologists using a qualitative scale with six levels. The diffusion and anisotropy coefficients (ADC and FA) were measured and the values compared with the data in the literature.ResultsMedian nerve tractography was successfully achieved in all sequences. The 20 direction sequence with b = 1000 s/mm2 seemed to provide the best results for a reasonable acquisition time (5′38″). The diffusion values showed large interindividual variations.ConclusionOptimisation of acquisitions parameters is important in peripheral nerve tractography, and may facilitate a routine use of the technique. The images obtained are easy to analyse but the diffusivity and anisotropy coefficients are too variable, however, to be used in practice.

Related Topics
Health Sciences Medicine and Dentistry Health Informatics
Authors
, , , , , ,