The case of a 78 year-old woman who underwent a right lower lobectomy using a 35-French, left-sided, double-lumen endobronchial tube (DLET) is presented. Multiple adjustments were needed for the DLET's proper placement. At the end of surgery, sudden loss of tidal volume with a large air leak from the patient's mouth was noted. Fiberoptic bronchoscopic examination through the DLET was negative. Rupture of the tracheal cuff was suspected, and the DLET was replaced with a single-lumen tube. In the intensive care unit, the massive air leak from the mouth recurred during mechanical ventilation. Nasal fiberoptic bronchoscopic examination showed a longitudinal laceration of the membranous portion of the trachea extending from the subglottic area to the orifice of the right bronchus. Surgical repair of the tear was performed.