Article ID Journal Published Year Pages File Type
2764140 Journal of Clinical Anesthesia 2008 5 Pages PDF
Abstract

Study ObjectiveTo evaluate a new disposable syringe (Certofix PresSure Check [B Braun Melsungen AG, Melsungen, Germany]) that serves as a manometer.DesignProspective study.SettingUniversity hospital's cardiac surgery suite.Patients21 consecutive patients with continuous invasive monitoring of central venous and arterial pressures during cardiothoracic surgery.Interventions and MeasurementsThe Certofix PresSure Check syringe was connected to the side port of the central venous and arterial catheters.ResultsCentral venous pressures were correctly indicated as below 30 mmHg and arterial pressures as above 50 mmHg in all 21 patients.ConclusionsCertofix PresSure Check provides a unique way of testing whether a catheter is placed in a low or high pressure system. It can be used for identifying inadvertent arterial punctures.

Related Topics
Health Sciences Medicine and Dentistry Anesthesiology and Pain Medicine