Article ID Journal Published Year Pages File Type
2764295 Journal of Critical Care 2017 5 Pages PDF
Abstract

PurposeWe assessed the security and efficiency of intravenously injected agitated saline in conjunction with transthoracic echocardiography to identify recirculation in patients supported with a venovenous extracorporeal membrane oxygenation (VV ECMO) device.Materials and MethodsWe injected agitated saline 4 consecutive times separated by an interval of 5 minutes in 2 patients supported by VV ECMO. In both patients, the drainage cannula was placed in the left femoral vein, and the return cannula was placed in the right internal jugular vein. Echocardiography was performed during the injection and until the bubbles disappeared. The security of the method was assessed by evaluating the mechanical function of the ECMO and the efficiency of the oxygenator. The value of this method was assessed by visualizing the increase of inferior vena cava's echogenicity as well as by measuring the time required for this change to occur after the injection of agitated saline at different ECMO output levels.ResultsWe did not observe any change in ECMO, oxygenation function, or the hemodynamic status of patients after the 4 injections of agitated saline. The echogenicity of the inferior vena cava increased more rapidly as the ECMO's output increased. The recirculation phenomenon was noted even with low levels of ECMO output (<2 L/min).ConclusionsTransthoracic echocardiography in conjunction with agitated saline administration may be a safe and easily applicable method to evaluate a recirculation phenomenon in patients supported with VV ECMO.

Related Topics
Health Sciences Medicine and Dentistry Anesthesiology and Pain Medicine
Authors
, , , , , , , ,