Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
2871153 | The Annals of Thoracic Surgery | 2016 | 7 Pages |
Abstract
The Magna Ease valves demonstrated excellent durability and hydrodynamic performance after an equivalent of 25 years of simulated in vitro wear. All study valves successfully endured 1 billion cycles of simulated wear, 5 times longer than the standard requirement for a tissue valve as stipulated in ISO 5840.
Related Topics
Health Sciences
Medicine and Dentistry
Cardiology and Cardiovascular Medicine
Authors
Vrishank PhD, Ikechukwu BS, Michael BS, Lynn BS, Salvador BS, MBA, Ajit P. PhD,