Article ID Journal Published Year Pages File Type
2886382 Annals of Vascular Surgery 2012 7 Pages PDF
Abstract

BackgroundA number of surgical strategies and graft enhancements have been trialled to improve the performance of prosthetic grafts. Neointimal hyperplasia may, in part, be a normal cellular response to an abnormal (turbulent) flow environment. This first-in-many study assesses the safety and medium-term patency performance of a new graft designed to induce stable laminar flow through the distal anastomosis.MethodForty patients who required an infrainguinal bypass graft were recruited/registered from a number of centers in Belgium and The Netherlands. Thirty-nine received a Spiral Laminar Flow graft as part of a standard treatment protocol (23 above-the-knee and 16 below-the-knee bypasses). Kaplan–Meier analyses were used to calculate primary and secondary patency rates.ResultsThe 12-, 24-, and 30-month primary patency rates were 86%, 81%, and 81% for above-the-knee bypasses and 73%, 57%, and 57% for below-the-knee bypasses, respectively. In the case of secondary patency rates, numbers were unchanged for above-the-knee bypasses and were 86%, 64%, and 64%, respectively, for below-the-knee bypasses. There were no amputations in the study population.ConclusionThis first-in-man series shows potential for the idea of spiral flow–enhanced prosthetic grafts. As always, randomized studies are required to explore the role of different enhanced prosthetic grafts.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , ,