Article ID Journal Published Year Pages File Type
2906206 Chest 2008 5 Pages PDF
Abstract

BackgroundThermal ablation is one of the most commonly used modalities to treat central airway obstruction. Both laser and argon plasma coagulation (APC) have been reported to cause gas emboli and cardiac arrest. We sought to determine whether bronchoscopic ablation therapy can result in systemic gas emboli, correlate their presence with the rate of gas flow, and establish whether a zero-flow (ZF) modality would result in the significant reduction or elimination of emboli.MethodsCO2 laser delivered through a photonic bandgap fiber (PBF) and APC were applied in the trachea and mainstem bronchi of six anesthetized sheep at varying dosages and gas flow rates. Direct epicardial echocardiography was used to obtain a four-chamber view and detect gas emboli.ResultsThe presence of gas flow accompanying APC and the CO2 laser with forward flow correlated significantly with the appearance of gas bubbles in the atria. A definite dose response was observed between the gas flow rate and the number of bubbles seen. When the CO2 laser was delivered through a PBF with ZF to the trachea or bronchi, no bubbles were observed.ConclusionBronchoscopic thermal ablation therapy using gas flow is associated with gas emboli in a dose-dependent fashion. The use of the flexible PBF with ZF is not associated with the development of gas emboli. Further study is required to determine whether a clinically safe threshold of gas emboli exists, and the relationships among the pathologic depth of tissue destruction, gas flow, pulse duration, and the development of gas emboli.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , , , , , , , ,