Article ID Journal Published Year Pages File Type
2915027 European Journal of Vascular and Endovascular Surgery 2006 4 Pages PDF
Abstract

ObjectivesEndovascular repair of abdominal aortic aneurysm is a technology that has undergone rapid technological evolution with a number of different types of grafts developed and concern has been the structural integrity of the metallic endoskeleton. We describe our experiences of strut failure in the Zenith abdominal aneurysm endograft device.Design/materials/methodsEighty-four patients have undergone endovascular repair of their abdominal aortic aneurysm at Christchurch Hospital from 1996 to 2005, all with the Zenith endoprosthesis. All available plain radiographs of the endoprosthesis (AP and lateral planes) were reviewed retrospectively, by a single experience observer, to assess strut failure.ResultsThree cases of strut failure in second-generation grafts were identified, each in the inferior body of the graft above the iliac bifurcation. There was no clear evidence that these strut failures were associated with clinical complications.ConclusionThis is the first time that strut failure in the Zenith abdominal endoprosthesis has been reported. Whilst in each of our three cases this does not appear to have been of clinical significance, these findings re-iterate the necessity of life-long surveillance of this technology.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine