Article ID Journal Published Year Pages File Type
2918428 Heart, Lung and Circulation 2014 4 Pages PDF
Abstract

The key to safe placement of a bicaval double lumen cannula for Venovenous Extracorporeal Membrane Oxygenation (VV ECMO) is to visualise correct guide wire placement in the inferior vena cava (IVC), thus aiding subsequent correct advancement of the cannula. Transoesophageal (TOE) and transthoracic (TTE) echocardiography, as well as fluoroscopy, have been described as aiding imaging techniques. We report a case of guide wire malposition into the right ventricle, despite echocardiographic confirmation of guide wire position deep into the IVC. This malposition, if undetected, may have resulted in potential life threatening complications.

Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine