BackgroundNavigation, mapping, and ablation in the right ventricular outflow tract (RVOT) can be difficult. Catheter navigation using external magnetic fields may allow more accurate mapping and ablation.ObjectivesThe purpose of this study was to assess the feasibility of RVOT tachycardia ablation using remote magnetic navigation.MethodsMapping and ablation were performed in eight patients with outflow tract ventricular arrhythmias. Tachycardia mapping was undertaken with a 64-polar basket catheter, followed by remote activation and pace-mapping using a magnetically enabled catheter. The area of interest was localized on the basket catheter in seven patients in whom an RVOT arrhythmia was identified. Remote navigation of the magnetic catheter to this area was followed by pace-mapping. Ablation was performed at the site of perfect pace-mapping, with earliest activation if possible.ResultsAcute success was achieved in all patients (median four applications). Median procedural time was 144 minutes, with 13.4 minutes of patient fluoroscopy time and 3.8 minutes of physician fluoroscopy time. No complications occurred. One recurrence occurred during follow-up (mean 366 days).ConclusionRVOT tachycardias can be mapped and ablated using remote magnetic navigation, initially guided by a basket catheter. Precise activation and pace-mapping are possible. Remote magnetic navigation permitted low fluoroscopy exposure for the physician. Long-term results are promising.