Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
295464 | NDT & E International | 2008 | 6 Pages |
Abstract
A new scanning technique for reducing artifacts due to the partial volume effect is presented which maintains the spatial resolution of the conventional method. These artifacts are stripes and they have a ring shape between the source and the rotation center. Experimental and simulation results are in good agreement. The scanning technique has two data acquisitions per one cross-section, “+” quarter-sensor offset and “−” quarter-sensor offset. The partial volume effect is reduced by mathematical averaging the corresponding data of two data sets and doing image reconstruction with the average data.
Keywords
Related Topics
Physical Sciences and Engineering
Engineering
Civil and Structural Engineering
Authors
Jun Nukaga, So Kitazawa, Hiroshi Kamimura,