Article ID Journal Published Year Pages File Type
2970305 The Journal of Heart and Lung Transplantation 2014 12 Pages PDF
Abstract
The 3 echocardiographic variables of measured changes in LV end-diastolic diameter, aortic valve opening time, and deceleration time of mitral inflow between the lowest (8,000 rpm) and highest pump speed settings (11,000 rpm) during echo-guided pump speed-change testing appear highly accurate in diagnosing device malfunction in the setting of pump thrombosis among patients supported with CF-LVAD. Further investigation is warranted to create and validate a prediction score.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine
Authors
, , , , , , , , , , , ,