Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
2988397 | Journal of Vascular Surgery | 2014 | 10 Pages |
Abstract
Distraction force acts against endovascular fixation to provoke stent graft migration. These blood flow simulations describe the relationship between morphologic features of the aortoiliac vessels and distraction force. A greater understanding of this relationship may help avoid endovascular solutions that carry a high risk of migration. In some cases, this may mean an alternative choice of seal zone or even device. In others, it could lend support to an open surgical approach or to a more intensive follow-up regimen.
Related Topics
Health Sciences
Medicine and Dentistry
Cardiology and Cardiovascular Medicine
Authors
Steven M. MBChB, Robert J. PhD, Thien V. PhD, Rachel L. PhD, Richard G. FRCR, John A. MD, FRCS, Srinivasa R. MD, FRCS, Robert K. MD, FRCS,