Abstract

SummaryThe efficiency of a pulsed biphasic waveform (PBW) was compared with that of biphasic truncated exponential (BTE) waveforms. First defibrillation shock outcome was studied in a population of 104 out-of-hospital cardiac arrest patients in ventricular fibrillation as the presenting rhythm. The call to first shock time was 8.2 ± 5.4 min. At 5 s post-shock, defibrillation efficiency was 90%. The arrest was witnessed in only 50% of the patients and only 5% received bystander CPR. Despite these limitations 38% of the patients achieved restoration of a spontaneous circulation at departure from scene and 9.8% were discharged from the hospital. These observations demonstrate a rate of first shock success in termination of ventricular fibrillation comparable to that reported with biphasic truncated exponential waveforms in out-of-hospital cardiac arrest.