3010703 | Resuscitation | 2007 | 7 Pages |
Abstract
The use of oxygen to drive the LUCAS device results in a rapid increase in ambient oxygen concentration to levels likely to risk injury or death from fire. Ambulance services and hospitals using the device must be alerted to these dangers immediately.