Article ID Journal Published Year Pages File Type
3010703 Resuscitation 2007 7 Pages PDF
Abstract
The use of oxygen to drive the LUCAS device results in a rapid increase in ambient oxygen concentration to levels likely to risk injury or death from fire. Ambulance services and hospitals using the device must be alerted to these dangers immediately.
Related Topics
Health Sciences Medicine and Dentistry Cardiology and Cardiovascular Medicine