Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3096510 | World Neurosurgery | 2013 | 7 Pages |
ObjectiveTo evaluate implant accuracy and cosmetic outcome of a new intraoperative patient-specific cranioplasty method after convexity meningioma resection.MethodsThe patient's own bone flap served as a template to mold a negative form with the use of polymethyl methacrylate (PMMA). The area of bone invasion was determined and broadly excised under white light illumination with a safety margin of at least 1 cm. The definitive replica was cast within the remaining bone flap frame and the imprint. Clinical and radiologic follow-up examinations were performed 3 months after surgery.ResultsFour women and two men (mean age 51.4 years ± 12.8) underwent reconstruction of bone flap defects after meningioma resection. Mean duration of intraoperative reconstruction of the partial bone flap defects was 19 minutes ± 4 (range 14–24 minutes). Implant sizes ranged from 17–35 cm2 (mean size 22 cm2 ± 8). Radiologic and clinical follow-up examinations revealed excellent implant alignment and favorable cosmesis (visual analogue scale for cosmesis [VASC] = 97 ± 5) in all patients.ConclusionsPatient-specific reconstruction of partial bone flap defects after convexity meningioma resection using the presented intraoperative PMMA cast method resulted in excellent bony alignment and a favorable cosmetic outcome. Relatively low costs and minimized operation time for adjustment and insertion of the cranioplasty implant justify use of this method in small bony defects as well.