Article ID Journal Published Year Pages File Type
3117285 American Journal of Orthodontics and Dentofacial Orthopedics 2011 8 Pages PDF
Abstract

IntroductionThis study evaluates the changes in the force system of the beta-titanium T-loop spring (TLS) caused by stress relaxation.MethodsNinety TLSs with dimensions of 6 × 10 mm, made of 0.017 × 0.025-in beta-titanium alloy and preactivated by concentrated bends, were randomly distributed into 9 groups according to the time point of evaluation. Group 1 was tested immediately after spring preactivation and stress relief by trial activation. The other 8 groups were tested after 24, 48, and 72 hours, and 1, 2, 4, 8, and 12 weeks. By using a moment transducer coupled to a digital extensometer indicator adapted to a universal testing machine, the amounts of horizontal forces and moments and the moment-to-force ratios were recorded at every 0.5 mm of deactivation from 5 mm of the initial activation in an interbracket distance of 23 mm.ResultsThe horizontal forces and moments were higher (P <0.001) for group 1 compared with the other 8 groups, which were not different among themselves. All groups produced similar moment-to-force ratios (P = 0.600), with no influence of time.ConclusionsThe TLSs preactivated by concentrated bends had progressive load decreases over time, and this effect is critical in the first 24 hours.

Related Topics
Health Sciences Medicine and Dentistry Dentistry, Oral Surgery and Medicine
Authors
, , , , ,